Copyright 2022 Baker Hughes Company.
200 | DPI610E Instruction Manual–English
Chapter 11. Documenting
11.7 Making an adjustment on the Device Under Test (DUT)
Step 1
If at the end of the test procedure, an
adjustment is required. For example, if the
end result of the test procedure is a
Fail
.
Then it is possible to perform and verify an
adjustment using the
Analysis
softkey.
Step 2
Make the necessary adjustment to the
DEVICE UNDER TEST (DUT) and check
the output signal through its entire
calibration range to ensure it is within spec
before repeating the calibration test
procedure.
When the adjustment is complete, select the
Back
softkey to return to the
Procedure Summary
screen.
Step 3
The test procedure may now be done again
after adjustment. Select the
Play
softkey or icon to do this.