40 | DPI 610E Instruction Manual–English
Chapter 3. Basic Tasks
3. Basic Tasks
3.1 Tasks
Use the Dashboard to access the
Tasks
application. The
Tasks
menu has a list of tests that
automatically configure the instrument when selected. The following options are available on the
Tasks
screen:
3.1.1 P-I (Pressure to Current measure)
This configures channel
CH1
to measure pressure, which can be internal (
INT
) or external (
EXT
).
If external pressure is detected on
CH1
, it will retain this function, otherwise it defaults to internal
pressure.
Channel
CH2
is configured to measure Current.
This task is typically for calibrating Current Output Pressure transmitters.
3.1.2 P-P (Pressure to Pressure)
This configures
CH1
to internal (
INT
) pressure measurement and
CH2
to external (
EXT
) pressure
measurement.
Step 1
Select the
Tasks
icon on the Dashboard
to show a list of available tasks.
Tap on the task to select it.
Note
: Any
Task
option added to the
Dashboard screen will be identified by a
mini
Home
icon next to the option text.
(As shown in the second screen).
Basic Tasks