Copyright 2022 Baker Hughes Company.
144 | DPI610E Instruction Manual–English
Chapter 8. External Sensors
8.1.4 The Zero Function
Note:
It is recommended that all gauge sensors are zeroed at the beginning of each day before
use.
8.1.5 Available External Pressure Sensors
Only the external pressure sensors in Table 8-2 are compatible with the DPI 610E type of
instrument.
Table 8-2: External Pressure Sensors for DPI 610E Range
INFORMATION
The Zero function is only available on gauge sensors. It is
not possible to apply total vacuum to absolute sensors, because they can only
measure atmospheric pressure. Refer to “The Zero Function” on page 75 for
more information about the use of this function.
Pressure Range
Gauge
Absolute
Differential
Barometric
25 mbar / 10 in H20/ 2.5 kPa
008G
-
008L
-
70 mbar / 1 psi / 7 kPa
01G
-
01L
-
200 m bar / 3 psi / 20 kPa
02G
-
02L
-
350 m bar / 5 psi / 35 kPa
03G
03A
03L
-
700 m bar / 10 psi / 70 kPa
04G
04A
04L
-
1 bar / 15 psi / 100 kPa
05G
05A
05L
750 - 1150 m bar / 11 - 17 psi /
75 - 115 kPa (Barometric)
-
-
-
05B
2 bar / 30 psi / 200 kPa
07G
07A
07L
-
3.5 bar / 50 psi / 350 kPa
08G
08A
-
-
7 bar / 100 psi / 700 kPa
10G
10A
-
-
10 bar / 150 psi / 1000 kPa
11G
11A
-
-
20 bar / 300 psi / 20 MPa
13G
13A
-
-
35 bar / 500 psi / 2 MPa
14G
14A
-
-
70 bar / 1000 psi / 7 MPa
16G
16A
-
-
100 bar / 1500 psi / 10 MPa
165G
165A
-
-
135 bar / 2000 psi / 13.5 MPa
17G
17A
-
-
200 bar / 3000 psi / 20 MPa
18G
18A
-
-
350 bar / 5000 psi / 35 MPa
-
20A
-
-
700 bar / 10 000 psi / 70 MPa
-
22A
-
-
1000 bar / 15 000 psi / 100 MPa
-
23A
-
-
1400 bar / 20 000 psi / 140 MPa
-
24A
-
-