132 | DPI610E Instruction Manual–English
Chapter 7. Utilities
Step 3
On the Leak test screen,
CH1
will
automatically be maximized to show the
relevant test details. The
Leak
icon
is displayed below the Function name
field. The
WAIT TIME
and
TEST TIME
are the two parameters for setting up
the Leak test and these are displayed in
the channel window. They use the
HH:MM:SS format.
To enter a value into either of the
parameters, tap the
WAIT TIME
or
TEST TIME
text to show an onscreen
keypad. Use this keypad to enter the
required value.
WAIT TIME
- Before starting a Leak test
a period of time may be required to
allow the pressurized system to settle
before the actual test is started. This
time is referred to as
WAIT TIME
on the
DPI 610E and the default
WAIT TIME
is
1 minute (00:01:00). This time value can
be changed to any value between
0 seconds (00:00:00) up to 60 minutes
(01:00:00).
TEST TIME
- This is the period over
which the DPI 610E is testing for
change in pressure i.e., leaks. The
default
TEST TIME
is 1 minute (00:01:00) and
this time value can be changed to any
value between 1 second (00:00:01) and
480 minutes (08:00:00).
These time settings can be quickly
accessed if using the touchscreen, by
tapping the relevant fields on the
channel window. Another way to access
these settings is by selecting the
Setup
softkey.
The
Leak Test Setup
screen will
provide the options for both the
WAIT TIME
and the
TEST TIME
.
Press the
Back
softkey to return to the
Leak Test
screen.
Step 4
After the Leak Test times have been set, pressurize the system to the desired pressure using the
DPI 610E pump.