102 | DPI610E Instruction Manual–English
Chapter 6. Calibrator Tasks
5.
RAMP
The
RAMP
function allows the source value to be automatically incremented from a defined
START
value to a defined
END
value in steps of a defined value in both increasing and
decreasing direction. The
TRAVEL
time can be set to define the period of time it takes the value
to go from
START
to
END
(and vice versa).
START
value - is the first current source value to be set using this option.
END
value - is the maximum current source value set-point using this option.
Step 1
From the
Source Automation
screen
tap the
DEFINED STEP
radio
button to
show the
Defined Step Automation
screen.
If required, select the
Edit
softkey to
change the selected automation values.
Click the
Proceed
icon to activate
the option and return to the
Calibrate
screen.
Step 2
Tap or select the parameter value to be
changed.
Use the onscreen keypad to change the
value.
For a description of
Scaling
refer to
Select the
Tick
icon
to set the new
values.
Step 3
The
Calibrator
screen will have two
additional icons indicating that the
Ramp
mode is active and the
Auto-repeat
operation is available.