PERFORM CALIBRATION
Step 7
Enter the
REFERENCE VALUE 1
value as
displayed on the external calibrated
equipment. Press the Tick
softkey to
confirm and return to the
Calibration
screen.
Step 8
Press the
Next
softkey to complete
Calibration point 1
and proceed to
Calibration point 2
.
Step 9
Apply Steps 6 to 8, but this time begin by
applying the nominal zero Current
(approximately 0 mA) to the DPI 610E. This
completes Calibration point 2.