PERFORM CALIBRATION
To calibrate an electrical function (Current Measure is used in this example), make sure the
electrical connection between the DPI 610E and the external calibrated equipment is correct
(See Section 6.2.9 on page 88 for Current Measure electrical connection diagram).
Step 1
Select
PERFORM CALIBRATION
from the
Instrument Calibration
menu.
Step 2
Select a
FUNCTION
option (e.g.,
Current
).
Step 3
Select a
DIRECTION
option if applicable
(e.g.,
Measure
).