Functions
Step 3
Select
Millivolts
>
Measure
then select
one of the following:
Off
for measurement without internal
loop power supply OR
10 V
for measure with internal 10 V loop
power supply (only for non-IS units) OR
24 V
for measure with internal 24 V loop
power supply.
Select the Tick
icon to confirm the
selection and configure the instrument.
Check that Voltage mode is selected.
Select the
Back
icon to show the
Calibration
screen.
Check that Millivolts is shown in screen.