English–DPI610E Instruction Manual | 145
PM700E
8.1.6 Configuration of an External Pressure Sensor
To configure a channel with the external remote pressure reading:
The
EXT
ernal pressure function is similar to that of the
INT
ernal pressure function because it
supports all the Calibrator task features associated with that of the internal pressure e.g. sensor
zero, pseudo-range, leak test, switch test, relief valve etc.
Step 1
Select the desired channel (channel 1 in
this example).
Step 2
Select
FUNCTION
from the Channel
Setup
screen.
Step 3
Select
Pressure
>
EXT
>
Normal
or
Pseudo
.
Select the
Tick
icon to confirm the
selections and show the
Channel
Setup
screen.
Refer to “Normal and Pseudo Pressure
Range” on page 77 for more information
about the use of Normal and Pseudo
sensor values.