Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual| 93
Functions
Direct Input of Current Value
You can directly input the current value to be sourced. First select the type of test to be done (See
Section 3.2 on page 42). The Touchscreen or the Navigation Pad can then be used to input this
current value:
The DPI 610E can provide electrical current
accurately within the 0 to 24 mA range.
The diagram on the left shows the terminal
connections for this configuration.
Use the Touchscreen to set the source
current
Tap in the
Current Value
field. Enter the required
current source value using the
UP
/
DOWN
arrow
icons on the left of the field.
Use the Navigation Pad to set the source
current
Press the
DOWN
button to select the
CH2
window area.
Press the
RIGHT
button until the source value is
outlined in bold.
Press the
OK
(
Enter
) button to show the
source value entry screen.Enter the new value
using the onscreen keyboard.and press the
Tick