Example Testing Session
11.5 Example Testing Session
Step 1
Using the step instructions in the title bar, go
to (or input) the Step 1 setpoint value
displayed.
In this example, apply 0.0000 bar using the
DPI 610E pump and/or volume adjuster.
When the value is within the test point
tolerance limits, select the
Add Test Point
softkey to capture the datapoint.
Step 2
A popup message window appears which
shows the
Pass/Fail
status of both the input
(test point) and the output (result) and
requires confirmation to log the calibration
data point. Select
OK
to accept and log the
reading or
Cancel
to abort.
Step 3
Go to the next step setpoint as shown in the
title bar and repeat the previous step.
Repeat until all steps have been completed
and logged.