INTERNAL PROCEDURES
Step 4
This is the test procedure setup screen
which has the following sections which are
critical to the procedure:
1. Input.
2. Output.
3. Test Points.
4. Tolerance.
These details are auto populated based on
the current Calibrator configuration,
parameters configured in the Analysis
application setup and other default settings.
Details of each section is given the
following steps.
INPUT
: The input refers to the test input
signal to the Device Under Test (DUT). The
input function type, the sensor type (where
applicable), the range, and the
measurement units all appear in this
section.
By default, the Function (and hence Sensor
type) set up in
CH1
is configured as the
input with the selected
Units
. These are
read-only and not editable in the procedure
creation wizard. If any changes are
required, these must be done in Calibrator
prior to using the Documenting application.
The input range is auto populated by
default with the full sensor range related to
the Function. Only the range can be edited
and must be within the full sensor range.
Set the
START
and
END
values of the input
range if different to the values displayed.
Note
: It is possible using the Toggle softkey
to swap the Input and Output functions i.e.,
use the original Input function as the Output
function and vice versa.