English–DPI610E Instruction Manual | 163
Periodic Trigger Options
9.3.3 DATA POINTS
When a Periodic trigger is selected, the number of data points configured is displayed in the
Data
Log Setup
screen. The number of data points is directly linked to the time interval and the log
duration configured. For example, if a time interval of 10 seconds (00:00:10.0) and log duration
of 1 minute (000:01:00) is configured, the number of data points displayed in the setup menu will
be 6. Each time the time interval or log duration value is changed, the number of data points is
automatically adjusted. Alternatively, the periodic data log can be configured by specifying the
number of data points required and the time interval or sampling frequency. From the previous
example, if the number of data points is changed from 6 to 5, the log duration will automatically
adjust to 50 seconds (000:00:50) based on the unchanged time interval of 10 seconds and the
new number of required data points.
9.3.3.1 How to set the DATA POINTS
Step 3
Check that required duration appears in the
LOG DURATION
section of the
Data Log
Setup
screen.
Step 1
Select
DATA POINTS
from the
Data
Log Setup
screen