English–DPI610E Instruction Manual | 177
Analysis Function
10.3.3 ERROR TYPE
The error or deviation from the transfer characteristic can be calculated and expressed in any of
the following options:
The default selection is %Span.
10.3.4 TOLERANCE
The tolerance or test limit values for the calculated result error or deviation from the transfer
characteristic, can be defined using this option. The tolerance value is expressed either as a %
or as an absolute or fixed measurement unit e.g., mA. This is reliant on the Error type selected.
The default tolerance value is 0.1%.
10.4 Analysis Function
Set the
Input
and
Output
channel and error analysis parameters as described in “Analysis
Select the Play
button to start.
The
Analysis
main screen window displays the following:
1.
The deviation tolerance type.
2.
The error/deviation value of the output channel from the ideal transfer characteristic.
3.
A live indication of the tolerance result status icon which is one of the following:
PASS
- the currently measured output value lies within the specified tolerance limits.
FAIL
- the currently measured output value lies outside the specified tolerance limits.
% Span
- a percentage of the Output signal span.
% Full Scale
- a percentage of the Output signal full-scale.
% Reading
- a percentage of the Output signal reading.
Fixed
- absolute measurement units of the Output signal.
1
2 3