88 | DPI610E Instruction Manual–English
Chapter 6. Calibrator Tasks
6.2.9 Current
The current measurement range: +/- 55 mA.
The DPI 610E can measure or source electrical current expressed in milliamps (mA). Only
CH2
can do this. When using the
CH2 Current
function, there is the additional option to use the
internal 10 V (non-IS instruments only) or 24 V power supply provided by DPI 610E or to use an
external power supply for the device under test.
6.2.9.1 Current Measure
It is necessary to use the correct terminals to configure the
Current Measure
function:
Step 1
Use the diagram on the left to make the
necessary connections for Current
Measure when the 10/24V internal
power supply is not used (the
Off
option
is selected in Step 4).
Use the diagram on the left to make the
necessary connections for Current
Measure using the internal 10 V or 24 V
supply.