Chapter 12. HART
12.2 HART Device Configuration
Step 1
When the
HART
device has been
physically connected to the DPI 610E
(electrical connection and/or pressure
connection), the loop current measured
by the DPI 610E is displayed as the
secondary reading in the
CH2
window.
To establish communication between
the instrument and the
HART
device,
select the
Configuration
softkey.
Note
: When the
HART
device has no
communication with the DPI 610E, an
animated set of dashed lines (-----)
appears in the primary reading area.
Step 2
Alternatively, the
Configuration
menu
can be accessed from the
CH2 Setup
screen.
Note
: if the
HART
application is already
running, maximize the
CH2
window to
access the
CH2 Setup
softkey.
Procedure continues with Step 3 on next page