Chapter 11. Documenting
Step 4
Cont’d
Linearity
– the relationship between the
input and output can also be specified i.e.,
either Linear or Square Root transfer
function.
The default is Linear.
TOLERANCE
– this is the maximum
deviation or permissible error margin for
each input test point in the calibration. It is
specified as a percentage of the Input
range.
The default test point tolerance value is 5%
and can be changed if required.
TOLERANCE
: This section refers to the
deviation on the output signal or result with
respect to each input signal setpoint
applied.
PASS/FAIL
(or
Result Tolerance
) (
P/F
):
This sets the maximum deviation that
determines if the result (output) at each test
point is within specification (
PASS
) or out of
specification (
FAIL
). It is expressed as a
% of the output or in fixed measurement
units depending on the tolerance type.
The default value is 0.1% (based on% FS).