Tasks
3.1.3 P-V (Pressure to Voltage)
This configures
CH1
to measure pressure, which can be internal (
INT
) or external (
EXT
). If
external pressure is detected on
CH1
, it will retain this function otherwise, it defaults to internal
pressure.
CH2
is configured to measure Voltage.
This task is typically for calibrating Voltage Output Pressure transmitters.
3.1.4 I-P (Pressure to Current source)
This configures
CH1
to measure pressure, which can be internal (
INT
) or external (
EXT
). If
external pressure is detected on
CH1
, it will retain this function otherwise, it defaults to internal
pressure.
CH2
is configured to Current source.
This task is typically used for calibrating I/P pressure converters.
3.1.5 P-Display (Pressure to Display)
This configures
CH1
to measure pressure which can be internal (
INT
) or external (
EXT
). If
external pressure is detected on
CH1
, it will retain this function otherwise, it defaults to internal
pressure.
CH2
is configured to Observed function.
This task is typically used for calibrating pressure devices with no electrical output but which have
a visual indication of measured pressure.
3.1.6 Leak Test
This configures
CH1
to pressure measure which can be internal (
INT
) or external (
EXT
) with the
Leak Test
utility. If external pressure is detected on
CH1
, then the external pressure
configuration is kept, otherwise
CH1
defaults to internal pressure configuration.
CH2
configuration remains unchanged.
For more details on the
Leak Test
, refer to Section 7.1 on page 130.
3.1.7 Switch Test
This configures
CH1
to measure pressure, which can be internal (
INT
) or external (
EXT
) while
the
Switch Test
utility details are displayed on
CH2
. If external pressure is detected on
CH1
, it
keeps this function, otherwise it defaults to internal pressure.
For more details on the
Switch Test
, refer to Section 7.2 on page 135.
3.1.8 TX SIM (Transmitter Simulation)
This configures
CH1
to measure pressure, which can be internal (
INT
) or external (
EXT
). If
external pressure is detected on
CH1
, it will retain this function otherwise, it defaults to internal
pressure.
CH2
is configured with Current source (in transmitter simulation mode).
For more details on the
TX SIM
task, refer to Section 7.3 on page 138.
3.1.9 Relief Valve Test
This configures
CH1
to measure pressure, which can be internal (
INT
) or external (
EXT
) with the
Relief Valve Test
utility. If external pressure is detected on
CH1
, it will retain this function
otherwise, it defaults to internal pressure.
CH2
configuration remains unchanged.
For more details on the
Relief Valve Test
, refer to Section 7.4 on page 140.