76 | DPI610E Instruction Manual–English
Chapter 6. Calibrator Tasks
To zero a sensor
Step 1
See Chapter 2.1.3 on page 27 for pneumatic units or
Chapter 2.2.4 on page 39 for hydraulic units.
Make sure that the sensor is open to
atmosphere.
Example
: For the DPI 610E internal
sensor, open the pressure release knob
fully or make sure that the test port is
open to atmosphere.
Step 2
See Chapter 6.2.3.1 (INT) or Chapter 6.2.3.2 (EXT).
Select the desired pressure function
(
INT
or
EXT
) from the
CH Setup
menu
on
CH1
or
CH2
.
Step 3
Expand the pressure channel by
tapping anywhere in the channel
window (except the Units area) (See
Chapter 6.1.1.1, “Maximize and
Minimize Channel Window,” on
page 64).
Step 4
Select
to set the pressure sensor to
zero.