Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual| 73
Functions
6.2.2 None
Users can select this function option if they do not wish to display any function or readings on the
channel being setup. On the Calibrator main reading screen, the channel window will be cleared
of all readings and information with only the function name present.
6.2.3 Pressure
6.2.3.1 Internal Pressure
Pneumatic units contain internal pressure sensors which range from 350 mbarg to 35barg.
Hydraulic units contain internal pressure sensors which range from 70 bara/g to 1000bara.
Available internal pressure sensors are listed in Table 6-2.
Table 6-2: Internal Pressure Sensors in DPI 610E Range
Pressure
Pressure Range Code
Pneumatic
Hydraulic
350 mbar / 5 psi / 35 kPa
03
G
-
1 bar / 15 psi / 100 kPa
05
G
-
2 bar / 30 psi / 200 kPa
07
G
-
3.5 bar / 50 psi / 350 kPa
08
G
-
7 bar / 100 psi / 700 kPa
10
G
-
10 bar / 150 psi / 1000 kPa
11
G
-
20 bar / 300 psi / 2 MPa
13
G
-
35 bar / 500 psi / 3.5 MPa
14
G
-
70 bar / 1000 psi / 7 MPa
16
-
G or A
100 bar / 1500 psi / 10 MPa
165
-
G or A
135 bar / 2000 psi / 13.5 MPa
17
-
G or A
200 bar / 3000 psi / 20 MPa
18
-
G or A
350 bar / 5000 psi / 35 MPa
20
-
A
700 bar / 10000 psi / 70 MPa
22
-
A
1000 bar / 15000 psi / 100 MPa
23
-
A