Copyright 2022Baker Hughes Company.
English–DPI610E Instruction Manual | 211
HART Device Configuration
Step 3
The connected
HART
device can be
configured with its unique Poll Address.
The configuration screen provides
options for the
HART
device polling.
Select one of the following Device Poll
methods to use:
The connected HART device can be
configured with its unique Poll Address.
The configuration screen provides
options for the HART device polling.
Select one of the following Device Poll
methods to use:
1.
USE ADDRESS 0 ONLY
-
Searches for any connected
devices with Poll Address 0 only.
2.
FIND FIRST DEVICE
– Searches
the full Poll Address range
(0 to 63) and identifies the first
device found.
3.
CHOOSE POLL ADDRESS
–
Searches a single user specified
Poll Address.
4.
DEFINE CUSTOM RANGE
–
Searches a user specified Poll
Address range.
When the
CHOOSE POLL ADDRESS
is selected, the Poll Address number to
search must be entered.
The default is 0.
To edit the Address value, select the
Edit
softkey and enter the desired
value. Press the
Tick
softkey to
accept and the
Back
softkey to
return to the
Device Poll
screen.
Note
: Valid Poll Address value must be
between 0 and 63).
Step 3 procedure continues on next page