Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual| 103
Functions
TRAVEL
- time required for the current source value to ramp from the Start value to the End
value. This automatically calculates the required step size to be used to achieve the travel time.
DWELL
- time interval which can be set to pause or wait at the end of each direction or travel
before changing to the opposite direction of travel.
AUTO-REPEAT
- the tickbox option which allows the automation cycle to be repeated indefinitely
until manually stopped.
When selected, the
Ramp
icon is displayed at the top right of the
CH2
window on the
Calibrator
screen.
The Ramp sequence is not manual and hence the
UP
and
DOWN
nudge buttons are not usable
with this option. It is automated by pressing the automation
Play
softkey.
If selected in the configuration, the
Auto-repeat
icon is also displayed next to the
Ramp
automation
icon. The automation cycle is automatically repeated until manually stopped.
To stop the automation at any time, press the
Cancel
softkey.