Chapter 6. Calibrator Tasks
6.3 Process Options
6.3.1 Tare
The Tare function is used to set the channel readings to a temporary zero. When Tare is enabled,
the primary reading value is deducted from subsequent reading values until Tare is disabled. The
initial reading will be approximately zero.
For example, if the reading from the internal pressure sensor is 21.4985 bar at the time Tare is
enabled, the reading becomes approximately 0.000 as the value 21.4985 is captured and
subtracted from true reading and the resulting value is then displayed. Whenever Tare is enabled,
the Tare symbol will be displayed on the relevant channel window to indicate it is in operation.
Note:
Tare
is only available for
Measure
functions with the exception of
Barometer
and
HART
.
To enable the Tare function:.
Step 1
Select the desired channel.
Step 2
Select
PROCESS
.