PERFORM CALIBRATION
13.2.2 Calibration - Internal Pressure Sensor
To calibrate the internal pressure sensor of the DPI 610E, make sure the correct pressure
connection is made from the test port and the external calibrated pressure source. This sensor
calibration should only be done by approved service centres and personnel.
Step 13
The
Verification
mode shows live
reference and measured (or sourced)
values.
Here, different values or points within the
measured (or sourced) range can be
checked to verify the adjustment is
satisfactory.
Once verification is complete, select the
Back
icon to exit the calibration
procedure or select the
Restart
softkey
to repeat the calibration.
Step 1
Select
PERFORM CALIBRATION
from the
Instrument Calibration
menu.
Step 2
Select
Pressure
function.
Select
to confirm.