English–DPI610E Instruction Manual | 223
HART Service Methods
Step 6
Enter the required High Trim value and
select the
Tick
softkey to confirm.
Step 7
The High Trim value has been applied
and the measured value is now
displayed.
Note
: the signal measured by the
DPI 610E is denoted by the icon
,
while that measured by the HART
device is denoted by
.
Make sure that the reading is stable.
To perform the trim at this point, select
the
Trim
softkey.
Step 8
Select
OK
in the popup message
window to accept the Trim.
Or select
CANCEL
to return to the
previous step.
The High Trim is complete which
finishes this stage of the D/A trim
procedure.
The final step of the procedure is to
verify that the D/A trim adjustment is
sufficient.
Procedure continues on next page