English–DPI610E Instruction Manual| 101
Functions
LOW
value - is the first current source value to be set using this option.
HIGH
value - is the maximum current source value set-point using this option.
STEP SIZE
value - is the magnitude of each step increment or decrement. When using this
option, it is expressed as mA.
DWELL
- time interval which can be set to pause or wait at each set-point value before changing
to the next value.
AUTO-REPEAT
- the tickbox option which allows the automation cycle to be repeated indefinitely
until manually stopped.
When selected, the
Defined Step
icon is displayed at the top right of the
CH2
window on the
Calibrator screen.
The
Defined Step
sequence can be run manually by using the
UP
and
DOWN
nudge buttons to
increment and decrement the source values. It can also be automated by pressing the
automation
Play
softkey.
If selected in the configuration, the
Auto-repeat
icon is also displayed next to the
Defined
Step
Automation
icon. The automation cycle is automatically repeated until manually
stopped.
To stop the automation at any time, press the
Cancel
softkey.
Step 2
Tap or select the parameter value to be
changed.
Select the parameter row to show an
onscreen keypad, enter the new value
using the onscreen or Navigator keys.
Tap the
Back
icon to enter the
value and return to the
Calibrator
screen.
For a description of
Scaling
refer to
Step 3
The
Calibrator
screen will have two
additional icons indicating that the
Defined Step
mode is active and the
Auto-repeat
operation is available.