Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual | 251
EXTERNAL PRESSURE SENSOR CALIBRATION MENU
13.8 EXTERNAL PRESSURE SENSOR CALIBRATION MENU
See Section 13.1 on page 227 for how to access the
Calibration
menu from the Dashboard.
13.8.1 PERFORM CALIBRATION
To calibrate the external pressure sensor PM700E using the DPI 610E, MAKE sure the correct
pressure connection is made from the sensor and the external calibrated pressure source. Also,
ensure the sensor is connected to the
EXT SENSOR
port on the DPI 610E using the sensor cable
supplied. This sensor calibration should only be done by approved service centres and
personnel.
Step 1
Select
EXTERNAL PRESSURE
SENSOR
from the main
Calibration
screen.
Note
: To see the
External Pressure
Sensor
option, the
EXT
Pressure
function must be already configured in
the
Calibrator
menu and the sensor
connected successfully. See Section 8
on page 142.
Step 1
Select
PERFORM CALIBRATION
from
the
External Sensor Calibration
menu
screen.