Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual | 239
PERFORM CALIBRATION
13.2.3 Calibration – Internal Barometer
To calibrate the internal barometer sensor of the DPI 610E, make sure the correct pressure
connection is made from the barometer port and the external calibrated pressure source. This
sensor calibration should only be done by approved service centres and personnel.
Step 1
Select
PERFORM CALIBRATION
from the
Instrument Calibration
menu.
Step 2
Select
Barometer
function.
Select
to confirm.