Copyright 2022 Baker Hughes Company.
256 | DPI610E Instruction Manual–English
Chapter 13. Instrument Calibration
13.8.3.3 Change CALIBRATION DUE Date
Step 1
Select
CALIBRATION DUE
from the
Instrument Calibration
Date & Interval
screen.
Step 2
Select the required day, month and year, as
required (see Chapter 1.19.3 on page 25).
To change the value, select the row of the
variable. Press the value on the left of the
current value to decrease the value and the
value on the right to increase the value.
Repeat pressing of either left or right value
decreases or increases the selected current
value.
Select
to confirm the changes.