Chapter 6. Calibrator Tasks
How to configure an External Sensor as a Sealed Gauge
The
EXT
Pressure function must be configured in one of the channels and you need to go into
Calibrator to trigger this popup message.
Step 1
Switch on the instrument and select
the
Calibrator
icon and when the
popup window appears select
YES
.
If the instrument is already switched
on select the
CH1
or
CH2
channel.
INFORMATION
If NO is selected, the sensor will be used in its native form -
an Absolute pressure sensor.
Step 2
Make sure that the pressure port is
open to atmosphere and proceed
when the pressure reading is stable.
Select
to confirm.
Step 3
Make sure that
Sealed Gauge
is
correctly configured on the desired
channel.