Chapter 13. Instrument Calibration
Step 8
If
OK
is selected, the
CALIBRATION
COMPLETE
message appears in the status
box, to confirm the adjustment has been
made.
At this stage, there are three icons for three
ways to proceed. These are:
VERIFICATION
softkey navigates to
the Verification screen (see Step 9).
RESTART
softkey allows the
calibration procedure to be restarted if a
repeat calibration is required.
BACK
softkey exits the calibration
procedure and returns the user to the
Perform Calibration
menu screen.
Step 9
The
Verification
mode shows live
reference and measured (or sourced)
values.
Here, different values or points within the
measured (or sourced) range can be
checked to verify the adjustment is
satisfactory.
Once verification is complete, select the
Back
button to exit the Calibration
procedure or select the
Restart
softkey
to repeat the calibration.