Chapter 10. Analysis
The screen is in two sections. Each section displays information for a channel and identified as
either the
Input
or
Output
channel.
To test the full range of the of the Device Under Test (DUT):
•
Step the input signal value through its range. In the example screen shot, step the internal
pressure generated by using the DPI 610E pump from the lower range of the DUT to the
full-scale pressure value.
•
At each setpoint step, check the Analysis status at the top of the screen for the deviation.
•
When the test is complete, exit the Analysis screen using the
Back
button.
10.4.1 DATA LOGGING within Analysis
Checking a DUT in the Analysis application can be documented using the Datalog. To use this
feature, select the
Data Log
softkey. For more information on
Data Log
page 156. When using
Data Log
in the Analysis application, only the
Key Press
trigger mode is
available.
•
Enter the required Data Log file name
•
Select the
Start
softkey to proceed
•
At each
Setpoint
step (or when desired), press the “Add Datapoint”
softkey to capture
the current analysis data displayed by the screen.
•
When the testing is complete, press the
Back
button to exit the Datalog and Analysis
application.
Datalog files can be accessed via the
Data Log
application (Refer to “Data Log” on page 156).