English–DPI610E Instruction Manual | 207
HART® Connection
.
Step 3
Select the
Page Down
softkey to view
the second page of
CH2
functions.
Step 4
Select
HART
in the
FUNCTION
column.
Choose the desired
DIRECTION
(
Measure
or
Master
).
Select the loop power option (
Off
to use an
external power supply or
24 V
to use the
DPI 610E internal power supply).
Select the
Tick
icon to confirm the
selection and configure the instrument.
Step 5
In the
CH2 Setup
screen, the
HART
function is displayed.
The configuration status is also shown (i.e.
HART
device connected or disconnected).
See Section 12.2 on page 210 for
Configuration details.
The 250-ohm HART resistor option allows
the DPI 610E provide the resistor when
enabled (checkbox ticked).
If the instrument is connected directly to a
network, there must be a 250-ohm resistor
connected in series with the loop power
supply and the
HART
device. Here the
HART
resistor should be disabled in the
setup (checkbox unticked).