Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual | 193
The Documenting Main Screen
11.3 The Documenting Main Screen
This section describes the different elements of the
Documenting Main
Screen.
The title bar contains information about:
•
Number of steps the test requires and
what the current step is e.g., Step 1 of 2.
•
The input test point value to apply e.g.,
Go to 0.000 bar
•
Completion status at the end of the test.
The sensor information and readings for Input and
Output. The top channel displays the Input
information and the bottom channel, the Output.
In the output section, the Error Type is displayed
below the reading on the right side of the window.
Error and tolerance indicator is displayed for both the
input and the output.
The specified tolerance value is displayed on either
side of the zero mark. This value is based on the
Error Type and hence is expressed as a % unless the
Fixed units Error type is selected in which case is
expressed as the Output measurement units.
The calculated live error is displayed below the zero
mark both input and output channels. A marker is
also displayed to indicate the approximate error
value point with respect to the specified tolerance
and tolerance line.
Note
: if the marker is not visible, check the error
value as it most likely means the error is too large to
be visible on the error and tolerance indicator.