Copyright 2022 Baker Hughes Company.
248 | DPI610E Instruction Manual–English
Chapter 13. Instrument Calibration
13.5 BACKUP CALIBRATION
The current calibration configuration can be saved in the form as a Backup file. This Backup file
can be used to restore the instrument’s calibration if necessary.
Step 1
Select
BACKUP CALIBRATION
from the
Instrument Calibration
Date & Interval
screen.
Note
: Only one user calibration
configuration can be backed up at any one
time.
Step 2
Select
OK
to confirm backup.
Step 3
Check that the
CALIBRATION BACKUP
notification appears. If the notification does
not appear, repeat steps 1 and 2.