Post Examination of Test Procedure Results
11.6 Post Examination of Test Procedure Results
Step 1
Note
: When a test procedure is done on a
new Asset for the first time, there will be no
As-Found
or
As-Left
results.
When no
As-Found
or
As-Left
results are
detected for a particular asset, the
calibration data will be automatically saved
as
As-Found
.
If
As-Found
or
As-Left
results file is
detected, then at the end of the test
procedure there is the option to save the
results as
As-Found
or
As-Left
. Selecting
As-Found
will overwrite any existing
As-Found
data. If no
As-Left
data is
detected, and save as
As-Left
is selected,
then a new
As-Left
results file is saved. If
As-Left
data file exists already, then the file
is overwritten if
As-Left
option is selected.
Step 2
On the
Test Procedure Summary
screen,
the option to re-run the test procedure using
the same test details and device under test
(DUT) details is provided by using the
Play
softkey.
If you wish to exit, use the
Back
softkey to
return to the
Internal Files Procedure
menu.