Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual| 75
Functions
6.2.3.2 External Pressure
External pressure sensors (PM700E) are available in the range 25 mbarg/d to 1400 bara.
Refer to “External Sensors” on page 142 for a list of available sensors, and information about how
to configure the DPI 610E to recognize and use external sensors and RTD probes.
6.2.3.3 The Zero Function
The zero function is used on gauge sensors to remove offset drifts and therefore maintain the
highest accuracy.
Note:
It is recommended that all gauge sensors are zeroed at the beginning of each day before
use.
INFORMATION
See Chapter 6.2.3.4, “Normal and Pseudo Pressure Range,”
Step 4
Check that the desired configuration
appears in the Channel
Setup
screen.
Select
to return to the
Calibrator
Main
screen.
Step 5
Check that
INT Pressure
appears on
top left corner of the desired channel
screen.
INFORMATION
The Zero function is only available on gauge sensors. It is
not possible to apply total vacuum to absolute sensors, as they are designed
to measure atmospheric pressure.