Chapter 6. Calibrator Tasks
Step 3
Select
FUNCTION
from the
Channel
Setup
menu.
Step 4
Select
Current
>
Source
> then select
one of the following:
10 V
for measure with internal 10 V
Loop Power supply (only for non-IS
units) OR
24 V
for measure with internal 24 V
Loop Power supply.
Select
to confirm the selection.
If
Off
is selected the electrical
connections will not be the same as for
when 24 V or 10 V is selected (see
Step 1).
This screen will have only “Current” as
the Function when
Off
is selected.
When
24V
is selected, “Current (24V)”
will appear.
When
10V
is selected,
“Current (10V)”
will appear.
Select the
Back
icon to go to the
Calibrator Main
screen