English–DPI610E Instruction Manual | 189
INTERNAL PROCEDURES
11.2.5.2 Environment and User ID data
Step 1
If required, enter additional data into this
screen.
AMBIENT TEMPTERATURE
– Enter
the ambient temperature value where
the test is being done. Temperature
units available are °C or °F. Use the
Toggle
softkey to switch
between these units. The unit
conversion will be done automatically:
20°C is the default ambient
temperature value. Any value entered
must be between -100 to +100 °C (-
148 to 212 °F).
AMBIENT PRESSURE
– Enter the
ambient pressure value (or pressure of
the day) where the test is being done.
Pressure units available are mbar, psi
or in Hg. Use the
Toggle
softkey
to switch between these units. The unit
conversion will be done automatically.
For DPI610E pneumatic variants, the
default ambient pressure value is
obtained from the internal barometer
sensor. On hydraulic variants, the
default ambient pressure value is
1013 mbar. Any value entered must be
between 800 to 1200 mbar (11 to 18
psi or 23 to 36 in Hg).
AMBIENT HUMIDITY
– Enter the
ambient humidity value where the test
is being done. The default value is
70%. Any value entered must be
between 0 and 100%.