240 | DPI610E Instruction Manual–English
Chapter 13. Instrument Calibration
Step 3
The
Calibration - Barometer
screen has
three sections.
The top two sections show:
•
The reference reading from external
calibrated equipment.
•
The measured (or sourced) reading
from the DPI 610E.
The bottom section gives the following
information:
1.
Function minimum range (or negative
full-scale) value.
2.
Function maximum range (or positive
full-scale) value
3.
Calibration procedure step status
(visual status boxes and text status)
4.
User instructions for each step of the
procedure.
To start the calibration, apply nominal
negative full-scale or zero pressure as per
the displayed instruction. This is
Calibration point 1
.
When the measured Pressure reading is
stable, select the
Next
icon to continue.
Note
: Barometer sensor calibration
requires two valid calibration points.
Step 4
Enter
REFERENCE VALUE 1
as displayed
on the external calibrated equipment. Press
the Tick
softkey to confirm and return to
the
Calibration
screen.