Chapter 14. File System
14.1.1 The Files system screen options
14.2 Calibration
When the
Calibration
folder is selected, any previously exported calibration files will be listed. If
no export has been done before, select the
Export
softkey to generate the calibration files
stored on the unit. There are three types of calibration files:
Factory Calibration
– this is the default calibration carried out on the instrument in the factory
prior to shipping. It is saved in persistent storage and cannot be accessed directly or deleted by
the user.
Current Calibration
– This is the calibration data used by the instrument at any moment in time.
If the instrument is new and has not been used, the Current calibration will be the Factory
calibration. Once any calibration adjustment has been done on the instrument, this new data will
replace the factory calibration data as the new Current calibration. Subsequent calibrations will
overwrite this user calibration data.
Backup Calibration
– if the
Backup
calibration feature is used (See Section 13.5 on page 248),
a copy of the Current Calibration data is saved as a Backup data file.
Figure 14-1: Calibration Files screen
Option
Description
Calibration
Export / View instrument calibration files
Data Log
View / Open / Delete data log files
Procedures
View / Delete test procedures, assets, and results
files
Leak Test
View saved leak test result files
Switch Test
View saved switch test result files
Relief Valve
View saved relief valve test result files
Favorites
Proposed feature: yet to be implemented
Error Log
Export / View error log files
Event Log
Export / View event log files