Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual | 153
RTD Probe and Interface
8.2.5 RTD Profiles
The full-scale value of the RTD probe is displayed as well as the RTD probe type. When the RTD
function is selected, the desired RTD profile might need to be configured by:
•
Tapping on the currently saved RTD profile field in the
Channel Setup
screen.
•
Using the keypad to move the selector icon to the
RTD Profile
row and pressing the
ENTER
button to show a list of the available profiles.
The default profile set is the “Default EU” option or PT100-PA-392, which is based on the
European standard (DIN/IEC 60751) for Class A Platinum RTDs. There is also a “Default US” or
PT100-PD-385, which is based on the American standard. The Default EU and Default US profile
options are pre-defined and cannot be edited or deleted.
However, up to eight Custom profiles can be created by populating the user profile files available
with the required coefficients, temperature range and label.