258 | DPI610E Instruction Manual–English
Chapter 13. Instrument Calibration
Step 2
This screen has three sections.
The top two sections cover:
•
The reference reading from
external calibrated equipment.
•
The measured (or sourced)
reading from the DPI 610E.
The bottom section provides the
following information:
1.
Function minimum range (or
negative full-scale) value.
2.
Function maximum range (or
positive full-scale) value
3.
Calibration procedure step status
(visual status boxes and text
status)
4.
User instructions for each step
of the procedure.
To start the calibration, apply nominal
negative full-scale or zero value as per
the displayed instruction. This is
Calibration point 1
.
When the measured temperature or
resistance reading is stable, select the
Next
icon to continue.
Note
: RTD sensor calibration requires
two valid calibration points.
Step 3
Enter
REFERENCE VALUE 1
as
displayed on the external calibrated
equipment. Press the
Tick
softkey
to confirm and return to the
Calibration
screen.