38 | DPI 610E Instruction Manual–English
Chapter 2. Pump Operations
Step 14
Turn the pressure release knob fully clockwise, hand tight.
CAUTION
Do not operate the priming pump. Only use the volume adjuster
wheel.
Step 15
Slowly turn the volume adjuster wheel 2-5 turns clockwise to remove
any trapped air.[If mineral oil is used as the pressure media, operate the
pump gently to avoid creating a mess.]
Note:
The user may notice air bubbles at the opening of the test
port.
Step 16
The instructions in this Step 16 relate only to the connection of a hose to the test port and
removal of air from this hose. If a test device is to be connected directly to the instrument (or to
the hose after air has been removed), go to Step 17.
Connect the hose to the test port and use the Step 10 to Step 15 procedures to remove air from
the hose.
Use Steps 10 and 11. Ignore Step 12. Use Steps 13 and 14.
Step 15: Use the pump to move fluid into the hose. Stop using the pump when air bubbles do
not appear at the hose outlet.
CAUTION
Do not allow movement of hose to rotate adapter as this may
cause internal damage to the instrument.
Step 17
Hold the test device in position in the test socket while fully turning the swivel connector
counter-clockwise until it is hand tight. A blanking plug can be attached to the hose end instead
of a test device.
Note:
As required, use adapters provided with the instrument or an AMC adapter(s)
and the corresponding seal(s).
Step 17
Use the priming pump to prime the system to a maximum of 10-20 bar.