Copyright 2022 Baker Hughes Company.
English–DPI610E Instruction Manual | 257
EXTERNAL RTD SENSOR CALIBRATION MENU
13.9 EXTERNAL RTD SENSOR CALIBRATION MENU
Refer to Section 8 on page 142 for how to configure the DPI 610E to recognize and use an
external RTD sensor. This is necessary to enable the DPI 610E user interface to show the RTD
sensor calibration options.
13.9.1 PERFORM CALIBRATION
To calibrate the external RTD sensor using the DPI 610E, make sure the correct connection is
made between the RTD & RTD-Interface and the external calibrated resistance/temperature
source. Also, make sure the RTD-Interface is connected to the
EXT SENSOR
port on the
DPI 610E using the sensor cable supplied. This sensor calibration should only be done by
approved service centres and personnel in accordance with the instructions in Section 8 on
page 142. To access the
External Pressure Sensor Status
screen. Follow the steps below:
Step 1
Select
EXTERNAL RTD SENSOR
from the
main
Calibration
menu.
Note
: To see the
External RTD Sensor
option, the RTD function must be already
configured in the Calibrator menu and the
sensor connected successfully. Use the
cross-reference given at the top of this page.
Step 1
Select
PERFORM CALIBRATION
from
the
External Sensor Calibration
screen.