Vacuum Insufficient. See Op Manual
Appendix A
Problem/Condition
The pressure did not reach the value necessary for pre-plasma within 12
minutes. Load may be out-gassing.
Action
F
Run an empty chamber cycle. If it completes successfully, the
problem is in the load.
F
Repackage the load and restart the sterilizer.
F
If the empty chamber cycle is unsuccessful, perform a leak back test.
F
Check vacuum pump and add oil if needed.
F
Check PM log and perform PM if needed.
F
Replace throttle valve.
F
Replace vacuum pump.
Vacuum Insufficient for Plasma. Call ASP
Technical Service.
Problem/Condition
For the second consecutive cycle, the vacuum did not reach 525 mTorr
in the first 3 minutes of the plasma stage. Load may be out-gassing.
Action
F
Run an empty chamber cycle. If it completes successfully, the
problem is in the load.
F
If the empty chamber cycle is unsuccessful, perform a leak back test.
F
Check the vacuum pump and add oil if needed
F
Check the PM log and perform PM if needed
F
Replace the throttle valve
F
Replace the vacuum pump
F
Replace vent valve.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-37
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......