Corrective Action
F
If no beep, check the wiring and connections.
F
Check the pneumatic board connector.
F
Replace the sensor board.
BLOCK DOOR CENTER
NEXT SENSOR W/ CANCEL
or with the door closed.
DOOR CENTER BLOCKED
NEXT SENSOR W/ CANCEL
Action
F
Use the blue override button on the pneumatic solenoid to move the
door to block the sensor. A beep occurs when the sensor is blocked.
F
Press
CANCEL
for the next sensor test.
Corrective Action
F
If no beep, check the wiring and connections.
F
Check the pneumatic board connector.
F
Adjust the over center sensor flag.
F
Replace the sensor board.
Cassette Sensor Tests
Display
BLOCK CASSETTE SENSOR
NEXT SENSOR W/ CANCEL
Action
F
Use a cassette to block the cassette sensor. A beep occurs when
sensor is blocked. Remove cassette.
Display
CASSETTE SENSOR BLOCKED
NEXT SENSOR W/ CANCEL
AutoTest
Sensor Tests
B-10
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Summary of Contents for STERRAD 100