OPEN DOOR TIME-OUT. CALL TECHNICAL
SERVICE.
There is no printed message for this condition.
Problem/Condition
More than 15 seconds has passed since clearing the top sensor and the
door has not opened.
Action
F
Test the door bottom sensor, the speed control, the air valves and the
vacuum in the chamber.
F
Check the vacuum switch.
F
Replace the door bottom sensor.
F
Check for loose connections at the pneumatic pump.
Operator cancellation. Repackage
load. Restart sterilizer.
Problem/Condition
F
Operator canceled cycle.
or
F
Operator canceled cycle during the first part of the vacuum stage.
Action
F
Repackage the load.
F
Restart the sterilizer.
System Messages
Using System Messages
3-18
STERRAD® 100 Sterilizer
ASP Service Guide rev D