High/Low Pressure Range Divider Voltage
CAUTION: Be sure to use the correct leads on Block 2 systems to
avoid shorting out the power supply.
1. Open the control enclosure and interrupt power to the enclosure
using the safety interlock. Remove the microprocessor board, and
insert the AutoTest board.
2. Remove the connector from the Baratron pressure transducer, and
attach the voltage divider adjustment tool to the harness connector.
3. Apply power to the enclosure using the safety interlock.
4. Perform the Analog Test.
5. At the completion of the test, interrupt power using the safety
interlock, remove the AutoTest board and install the system
microprocessor board. Remove the adjustment tool and reconnect
the cable to the pressure transducer. Apply power to enclosure using
the safety interlock. Close the enclosure.
User Subsystem Replacement
Procedures
Remove/Replace the Printer
1. Unplug the ribbon cable. Remove the paper roll.
2. Unscrew the two screws holding the printer to the frame.
3. Lift up on the printer to remove it.
4. Replace the printer by tipping it into place and replacing the screws.
5. Plug in the ribbon cable. Replace the paper roll.
User Interface and Control Subsystems
User Subsystem Replacement Procedures
6-22
STERRAD® 100 Sterilizer
ASP Service Guide rev D.
Summary of Contents for STERRAD 100
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......