F
Adjust the injector pump optics.
F
Perform the injector test.
F
Verify that the injector solenoid is operational by pressing the
activation button.
F
Realign the chute with the injector pump.
F
Replace the injector pump.
PM Interval due: L1
Problem/Condition
700-750 cycles have been counted since the last PM. PM 1 is due. Also,
there is no cassette in the sterilizer.
Action
F
Insert a cassette to continue the current cycle.
F
Perform the Level 1 PM as soon as possible.
PM Interval due: L2
Problem/Condition
1450-1499 cycles have been counted since the last PM. PM 2 is due.
Also, there is no cassette in the sterilizer.
Action
F
Insert a cassette to continue the current cycle.
F
Perform the Level 2 PM as soon as possible.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-23